Absolute calibration of photodetectors: photocurrent multiplication versus photocurrent subtractionстатья
Статья опубликована в высокорейтинговом журнале
Информация о цитировании статьи получена из
Web of Science,
Scopus
Статья опубликована в журнале из списка Web of Science и/или Scopus
Дата последнего поиска статьи во внешних источниках: 18 июля 2013 г.
Аннотация:We report testing of the new absolute method of photodetector calibration based on the difference-signal measurement for two-mode squeezed vacuum by comparison with the traditional absolute method based on coincidence counting. Using low-gain parametric downconversion, we have measured the quantum efficiency of a counting detector by both methods. The difference-signal method was adapted for the counting detectors by taking into account the dead-time effect.