Heterogeneous Thermal-Lens Immunoassay for Small Organic Compounds: Determination of 4-Aminophenolстатья
Статья опубликована в высокорейтинговом журнале
Информация о цитировании статьи получена из
Web of Science,
Scopus
Статья опубликована в журнале из списка Web of Science и/или Scopus
Дата последнего поиска статьи во внешних источниках: 18 июля 2013 г.
Аннотация:4-Aminophenol was selected as a model hapten for thermal-lens detection in a heterogeneous indirect immunoassay for small organic compounds. The assay is based on the competition of the free hapten and 4-aminophenol labeled with tetramethyl rhodamine isothiocyanate (TRITC) for rabbit anti-4-aminophenol antibodies (4-AP-GA-BSA) immobilized on a transparent poly(ethylene terephthalate) plate. The amount of the colored analyte was directly measured by thermal lensing at the plate surface. The developed method is specific for 4-aminophenol. The limit of detection is 2 x 10(-7) M, which corresponds to the absolute amount of 3 x 10(-11) mol of 4-aminophenol. The sensitivities of the developed thermal-lens and fluorescence-based measurements of the immunoassay under the same conditions are in good agreement with the expected and discussed theoretical sensitivities of photothermal and fluorescence measurements.