Chromatographic determination of silicon and phosphorus as molybdic heteropoly acids with preconcentrationстатья
Информация о цитировании статьи получена из
Web of Science,
Scopus
Статья опубликована в журнале из списка Web of Science и/или Scopus
Дата последнего поиска статьи во внешних источниках: 18 июля 2013 г.
Аннотация:The chromatographic behavior of molybdic heteropoly acids of silicon and phosphorus as ion pairs with tetrabutylammonium bromide was studied by ion-pair high-performance liquid chromatography on the C 18 reversed phase (UV detection at 310 nm). Heteropoly acids were preconcentrated as ion pairs with tetrabutylammonium bromide to increase the sensitivity of the determination. Optimal conditions were selected for the chromatographic determination of silicon and phosphorus in water in the presence of each other. Detection limits for silicon and phosphorus are (1.1 × 0.3) × 10 -3 and (6.7 × 1.2) × 10 -3 μg/mL, respectively. Calibration plots are linear in the concentration ranges 0.01-0.1 μg/mL (silicon) and 0.02-0.15 μg/mL (phosphorus). The procedure was used for the analysis of distilled water.