Flow injection amperometry as an alternative to potentiometry for solid contact ion-selective membrane-based electrodesстатьяИсследовательская статья
Статья опубликована в высокорейтинговом журнале
Информация о цитировании статьи получена из
Web of Science,
Scopus
Статья опубликована в журнале из списка Web of Science и/или Scopus
Дата последнего поиска статьи во внешних источниках: 29 апреля 2021 г.
Аннотация:In this work we propose a flow-injection amperometry at constant potential as a perspective alternative to classical potentiometry for solid contact ion-selective electrodes (SC-ISEs). The amperometric readout mode eliminates the problem of background drift and potential instability – common limitations of potentiometry for SC-ISEs. The selectivity in flow-injection and constant flow amperometric modes is similar to the potentiometric one. Moreover, flow injection amperometry in comparison to potentiometry for ion-selective electrodes provides a shorter response time, within 15-30 s, by two orders of magnitude decreased detection limit and 5-6 fold increased signal-to-noise ratio. The proposed approach is an attractive readout mode that opens new horizons for the improvement of performance characteristics of solid contact ISEs.