Solid-phase extraction of alkylphosphonic and O-alkylalkylphosphonic acids followed by HPLC separation using porous graphitic carbon sorbentsстатьяИсследовательская статья
Статья опубликована в высокорейтинговом журнале
Информация о цитировании статьи получена из
Web of Science,
Scopus
Статья опубликована в журнале из списка Web of Science и/или Scopus
Дата последнего поиска статьи во внешних источниках: 1 сентября 2021 г.
Аннотация:An HPLC separation of alkyl phosphonic acids on porous graphitic carbon adsorbent Hypercarb, based upon a step gradient of formic acid concentration in an aqueous mobile phase, was evaluated in this study. Analytes were detected by single quadrupole and triple quadrupole MS. Good separation was achieved for methyl phosphonic acid, ethylphosphonic acid, n-propylphosphonic acid, isopropylphosphonic acid, ethyl methylphosphonic acid, isopropyl methylphosphonic acid, isobutyl methylphosphonic acid, pinacolyl methylphosphonic acid. Solid-phase extraction of these analytes on Hypercarb was also proposed, desorption was carried out with aqueous ammonium formate and water-methanol mixtures. LODs were 0.1-0.2 ng mL-1 for these analytes without SPE; solid-phase extraction resulted in the decrease of LODs at least 100-fold. Accuracy was proven by analysis of spiked samples of river water and snow meltwater.