Voltage-height spectroscopy in the ex situ configuration of a scanning tunneling microscopeстатья
Информация о цитировании статьи получена из
Web of Science,
Scopus
Статья опубликована в журнале из списка Web of Science и/или Scopus
Дата последнего поиска статьи во внешних источниках: 18 июля 2013 г.
Аннотация:Basic regularities of electrochemical processes in the gap of an ex-situ scanning tunneling microscope in conditions of condensation of air moisture at the sample surface are considered on a qualitative level. A layer of condensed moisture is viewed as an electrolyte in a two-electrode cell. The depolarizers present in this layer may experience electrochemical conversions on the tip and in an area of the sample surface near the tip. As a result, the recorded "tunneling" current includes electrochemical constituents. Depending on the electrochemical processes in the gap, various dependences of the tip-sample distance on the current and applied voltage can be expected. For preliminary diagnostics of processes in the gap it is suggested to use voltage-height spectra, whose shape and characteristic heights are sensitive to the nature and location of redox active species. Experimental data for various films on conducting supports (quasi-two-dimensional adsorbed layers of hemin and peroxidase, electrodeposited nonstoichiometric tungsten oxides, doped tin dioxide, solid electrolyte with ionic conduction) are presented as an examples.